Troubleshooting

Unable to connect to the camera/spectrometer

If the application is unable to detect the camera check the following:

  • When using a GiGe camera, check that the IP of the camera matches your current subnet using the camera manufacturer software.

  • Some GiGe camera can be powered via Ethernet (POE) but only if the Ethernet card supports it. Check that your card support POE or that the camera has an external power supply.

  • Check that the manufacturer camera software was installed properly and that all the necessary dll are installed.

  • Do not connect to the camera using another software as it will block camera access.

No images can be acquired

If the application connects correctly to the camera but fails to acquire image, check that you have installed the SDK version recommended in section 1/ as we already had issues when running the software with different SDK versions.

You may also check in SpinView that the following settings were set correctly to the camera and that SpinView can acquire image correctly:

Setting Value
AnalogControl > Gain Auto Off
AnalogControl > Gamma Enable unchecked
AnalogControl > Sharpness Enable unchecked
AnalogControl > Blacklevel Enable unchecked
AnalogControl > Hue Enable unchecked
AnalogControl > Saturation Enable unchecked
AcquisitionControl > Acquisition Mode Continuous
AcquisitionControl > Exposure Auto Off
AcquisitionControl > Exposure Mode Timed
AcquisitionControl > Framerate Enable(d) unchecked
AcquisitionControl > HDR Enable unchecked
AcquisitionControl > Framerate Mode Off
ImageFormatControl > Pixel Format Mono16
ImageFormatControl > Reverse X checked
ImageFormatControl > Color Processing unchecked
AcquisitionControl > Trigger Mode Off